20090177150 | Low Friction Systems and Devices | July, 2009 | Sivebaek |
20050113791 | Zoned absorbent structures and process for producing same | May, 2005 | Neubauer et al. |
20030183683 | Method and associated system for specimen and sample chain of custody tracking | October, 2003 | Stewart |
20090247949 | Single-Use Syringe Assembly | October, 2009 | Buder et al. |
20070142784 | INTRAVENOUS CATHETER ENCHORING DEVICE | June, 2007 | Dikeman et al. |
20030055404 | Endoscopic rotary abraders | March, 2003 | Moutafis |
20020049405 | Device and method for mixing a two-part composition forming synthetic bone | April, 2002 | Deslauriers et al. |
20080269707 | Lateral Flow Device for Attachment to an Absorbent Article | October, 2008 | Song |
20090221960 | BALLOON TROCAR ADVANCED FIXATION | September, 2009 | Albrecht et al. |
20080269668 | Medical Microbubble Generation | October, 2008 | Keenan et al. |
20050240148 | Traction cutting balloon | October, 2005 | Cheves et al. |
[0001] This is a Continuation of application Ser. No. 09/957,861, filed on Sep. 21, 2001.
[0002] Not applicable.
[0003] Not applicable.
[0004] This invention described in this Continuation relates to a manually-operated syringe, having ergonomic advantages, to be used for injecting fluids during medical procedures. More particularly, an intent of this ergonomic syringe is its use for injecting radiographically opaque fluid contrast medium into the vascular system of a patient during angiographic procedures, for the purpose of enhancing visualization of the vascular system on angiograms or other radiograms.
[0005] The radiographically opaque contrast medium, also called dye, is injected into the vascular system typically through a hollow catheter which has been inserted into a patient's artery or vein. For example, the catheter for coronary angiography is usually inserted into the femoral artery or radial artery. The open end of the catheter through which the dye is introduced into the blood flow is guided through the vascular system to the target area by the operator, normally a physician. The dye is injected through the catheter into the blood vessels being evaluated at the time that the angiogram is recorded. Because the dye is opaque to the x-rays used in angiography, it enhances the contrast on the angiogram so as to better show the interior topography of the blood vessels into which the dye is injected. A minimum density of dye in the blood flow of the vessels being evaluated is required in order for a diagnostically-useful angiogram to result. Dye normally has a higher viscosity than water or blood, with a measured viscosity, in centipoise, of between 2 and 20, where the viscosity of water is 1.
[0006] Dye is normally injected from a manually-operated syringe into an attached manifold, and therefrom into an attached catheter, these three items being in fluid communication by direct connection or by hollow tubing. The syringe, manifold, tubing, and proximal end of the catheter are located outside the patient's body. The connections between the syringe, manifold, tubing, and catheter are made using a threaded connector, typically a Luer connector. The manifold is used as a means of connecting the syringe, catheter and sources of dye and sometimes saline solution. By manipulating valves, also commonly known as stopcocks, the operator can open and close channels to the dye, saline and catheter, enabling desired fluid flow.
[0007] For coronary angiography, between six and 12 milliliters of dye are normally injected per angiogram; this injection should normally take approximately two seconds. If the injection takes substantially longer than approximately two seconds, the density of the dye in the blood flow decreases, lowering the contrast of the angiogram, thereby reducing its usefulness since the interior topography of the vessels being evaluated will not stand out sufficiently from the background of the angiogram. A catheter used for injecting dye is often also called a diagnostic catheter. Multiple injections of dye are required during each procedure, since many different angiographic views are recorded.
[0008] Operators injecting the dye normally use syringes made of plastic or polypropylene, with the barrel of the syringe held with the index and third finger placed either in rings or against flanges formed as part of the barrel, and with the plunger of the syringe actuated by the thumb. In many instances, a ring is formed as part of the end of the plunger, so as to enable forward and backward motion in response to the thumb's direction of movement. This style of syringe has been adequate to administer dye with the sizes of catheters in common use, for example, those sized 6 French or larger. French Size increments are in intervals of 0.3 millimeters; for example, a 6 French catheter has an outside diameter of 2.0 millimeters, and a 7 French catheter has an outside diameter of 2.3 millimeters. Inside diameters vary from catheter to catheter; a 6 French diagnostic catheter can be expected to have an inside diameter of approximately 1.3 millimeters, and a 7 French diagnostic catheter an inside diameter of approximately 1.6 millimeters.
[0009] Smaller catheters have been introduced because of benefits associated with the smaller puncture hole required to insert the catheters into the body; as the size of the puncture hole decreases, the risk of puncture site complications and the time required for the patient to ambulate decreases. These recently-introduced smaller diagnostic catheters, in 4 French and 5 French sizes, have narrower lumens; a 4 French catheter can be expected to have an inside diameter of approximately 1.1 millimeters and a 5 French catheter an inside diameter of approximately 1.2 millimeters. The inside diameter of a 4 French catheter can therefore be about 15% smaller than that of a 6 French catheter and about 30% smaller than that of a 7 French catheter.
[0010] The smaller inside diameters of the 4 French and 5 French catheters make manual injection of the viscous dye more difficult compared with larger catheters. This difficulty is caused by the increased amount of injection force required to propel the dye through a smaller lumen diameter to approximate the same density of dye in the blood flow as is normally achieved with larger-diameter catheters. The degree of difficulty in maintaining a minimum density of dye in the blood flow has not normally been a problem with catheters sized 6 French or larger. As lumen sizes decrease, operator hand strength becomes a limiting factor in the proper administration of dye, even when two hands are used to inject. Where the operator's hand strength is not sufficient, for example after repeated injections, lower quality angiograms result and the operator's wrist, hand and fingers become fatigued and incur a high risk of disability due to either acute or repetitive motion injury. This in turn reduces the adoption rate of the smaller diagnostic catheters, delaying realization of the benefits of lower complication risk and earlier post-procedure ambulation.
[0011] Therefore a means of injecting dye through catheters sized 5 French or smaller is needed, which reduces stress on the operator's hand, wrist and fingers while providing a density of dye in the blood flow of the vessels being evaluated sufficient for diagnostic purposes.
[0012] One approach to alleviating this problem includes the same type of manifold and catheter normally used in angiographic procedures, but which substitutes an electromechanical injector for the manually-operated syringe. Examples of such injectors are described in U.S. Pat. Nos. 6,221,045, 5,383,858, 4,854,324, 4,677,980 and 4,006,736. These injectors were initially developed for purposes of injecting the larger volume of dye into the ventricles of the heart required for ventriculography, and are operated by inputting instructions and then actuating the device, which then automatically injects the pre-set amount of dye at the pre-set flow rate. Although useful and generally safe for ventriculography, these injectors' use in angiography incurs additional risk because of the smaller tolerances involved with injecting into the much smaller volume of a blood vessel, combined with possible unplanned variation in rate or volume of dye administration. Another important disadvantage of these injectors is that, in the event of unforeseen vessel interior topography, malfunction, or inputting improper settings when injecting into a blood vessel, a rupture or other damage could occur resulting in severe adverse health outcomes including death. An additional disadvantage is the very high cost of the injectors, both for acquiring the equipment, maintaining it during its useful life, and for procuring the disposable supplies required for its operation. A further disadvantage is the injector's large size and complexity of operation.
[0013] Manually-operated syringes have not substantially changed since the introduction of the 4-French and 5-French catheters and are similar in general form and construction to that shown in U.S. Design Patent 320,276. Syringes in common use for angiography purposes are generally made of a plastic or polypropylene, with the barrel of the syringe held with the index and third fingers placed either in rings or against flanges formed as part of the barrel, and with the plunger of the syringe actuated by the thumb or a small portion of the palm of the hand. In many instances, a ring is formed as part of the proximal end of the plunger, so as to enable forward and backward motion in response to the thumb's direction of movement. None of the syringes in current use feature plungers with handgrips having a large surface area over which to distribute the pressure across a large portion of the operator's hand; such pressure is caused by application of manual force required to perform the injection. Specialized means of comfortably using two hands during the injection is not typically provided. In addition, features for finger placement on currently-used syringes are not spaced to permit best application of force during injection. A threaded connection, for example a Luer connector, is provided at the nozzle of these syringes for removable connection to the manifold which in turn connects using a similar connector to the tubing and catheter. For most operators, these syringes are often proving inadequate for use with the smaller catheters because of insufficient dye density leading to low diagnostic quality of the angiogram, and in addition, because of stress and pressure on operators' hands, wrists and fingers even when two hands are used. These types of syringes are marketed by companies examples of which include Merit Medical Systems, Inc., Boston Scientific, and DeRoyal.
[0014] A manually-driven syringe is described by Saez et al in U.S. Pat. No. 4,925,449, whereby a comparison and modification to a then-existing syringe is described. Features of this syringe include: finger grips on two opposing sides of the barrel, each of which includes an opening to fit two fingers; a plunger with a thumbring on its head, which, in an embodiment, may be axially collapsible; the distance between the plunger head and the placement of the operator's fingers in the finger grips being lessened as a result of the shape and location of the fingergrips and the location of the plunger head when extended. Although this syringe may represent an improvement over the then-current syringes, it has several disadvantages. A disadvantage is the requirement that two fingers be placed in each of the fingergrips, which departs from the normal current expectation of operators and the design of syringes in common use for this purpose, where only the index finger instead of the index and third fingers held together, is placed in a finger grip. Furthermore, the head of the plunger is shown to be similar in size to the then-current syringe of which it purports to be an improvement; this small size is a significant disadvantage in that it places excessive pressure on a small portion of the palm of the operator's hand or thumb when performing the injection. Such pressure is caused by application of manual force required to perform the injection. This disadvantage is exacerbated as the force required for injecting into the 4 French and 5 French catheters is significantly greater than the force required for injecting into the size of catheters commonly in use at the time the Saez invention was patented. A further disadvantage is the noted lack of specialized means for permitting two-handed use, which therefore reduces the potential utility of the syringe by limiting the amount of force to be applied during injection to that which can be applied with only a single hand, and increases the amount of discomfort associated with the injection. The current invention represents an improvement over Saez, in that the position of the top fingergrip relative to the bottom fingergrip is offset to permit greater ergonomic advantage to the user; specifically, the current invention has the top fingergrip located forward of the bottom fingergrip. Saez shows the finger grips being symmetrically opposed in their relative positions and identical in shape, where the current invention shows not only the positional offset of the fingergrips, but their assymetrical shape as well.
[0015] A manually-operated, mechanically-assisted power syringe is described by Anwar in U.S. Pat. No. 6,030,368, whereby a syringe is connected in a levered apparatus providing mechanical advantage to the injection action. Although providing mechanical advantage and reducing physical stresses on operators, a disadvantage of the power syringe is its relatively large size and corresponding requirement for stable, horizontal space not normally available near the catheterization procedure table. Another disadvantage is its high cost relative to other manually-operated syringes in common use, caused by the power syringe's design, size and construction. Another disadvantage is the power syringe's mode of use, which requires the operator to press down on a lever, actuating the injection action, while the power syringe lies on a flat surface near the patient; this fundamentally different operation is unfamiliar to operators who may find gauging flow rate and volume during injection more difficult.
[0016] Montuori describes a manually-operated syringe in U.S. Pat. No. 1,757,809, in which the illustration, but not the specification, shows a curved handgrip. Although an improvement over the then-current syringes, a disadvantage of the Montuori invention is its relatively short length of its handgrip. The handgrip referenced in
[0017] Masano, in U.S. Pat. No. 5,221,348, describes a manually operated high pressure glue injector. The fingergrips described in the current invention provide an improvement over those described by Masano, since the position of the top fingergrip relative to the bottom fingergrip is offset to permit greater ergonomic advantage to the user; specifically, the top fingergrip is located forward of the bottom fingergrip. Further, the bottom fingergrip is angled, also to permit greater ergonomic advantage. Masano describes a first set of opposed finger grips which each accommodate two fingers, and a second set of opposed finger grips placed forward of the first set which each accommodate one finger; both sets are symmetrical in their relative positions and identical in shape. The structure described by Masano is different from that described in the current invention, which, because of the offset in the position of the fingergrips and the angle of the bottom fingergrip, provide an improvement over the prior art.
[0018] An object of this invention, an ergonomic syringe, is to increase comfort and reduce fatigue, strain and risk of disability for operators using syringes to inject angiographic dye during catheterization procedures, particularly when catheters sized 5 French and smaller are being used.
[0019] A further object of the invention is to make injection of fluids during medical procedures easier.
[0020] An additional object of the invention is to enable acquisition of angiograms having sufficient contrast for proper diagnostic and therapeutic use.
[0021] An additional object of the invention is to reduce the cost of each syringe.
[0022] An additional object of the invention is that its usage be similar to that of manually-operated syringes commonly used in catheterization procedures.
[0023] These and other objects are achieved with the ergonomic syringe described herein. In an embodiment, the ergonomic syringe is generally comprised of the following basic elements: a hollow barrel; a nozzle; a top fingergrip; a bottom fingergrip; a plunger; a handgrip.
[0024] The ergonomic syringe is typically used during catheterization procedures, where it is connected inline to a manifold, which in turn is connected to the catheter using tubing through which the dye is injected into the target blood vessels. The syringe, manifold, tubing and catheter are in fluid communication. The manifold, tubing and catheter are distinct elements separate from the syringe. Manipulation of the syringe by the operator causes the plunger to move longitudinally, in either a rearward or forward direction. Rearward plunger movement, towards the proximal end of the syringe, draws dye into the barrel from the manifold; forward plunger movement, towards the distal end of the syringe, expels dye into the manifold and therefrom into the catheter.
[0025] A hollow nozzle is located at the distal end of the barrel. The nozzle and barrel are in fluid communication, and the nozzle also may serve as a connector to a manifold and catheter, or other means of injecting dye.
[0026] Fingergrips are located on the exterior walls of the barrel, generally opposite each other, and include a top fingergrip and a bottom fingergrip.
[0027] The top fingergrip includes an open finger ring to contain an index finger, and an area which may be used to rest the index finger of the opposite hand, comprising an anterior top fingergrip and anterior top fingergrip catch. This top fingergrip design specifically enables, but does not require, use of two hands during dye injection. The bottom fingergrip includes an open interior bottom fingergrip, into which the third, fourth and little fingers of a hand may be inserted. The bottom fingergrip further includes an anterior bottom fingergrip around which the third, fourth and little fingers of the opposite hand may be placed, and a little finger rest which prevents the little finger of said hand from slipping off of the anterior bottom fingergrip. These features on this bottom fingergrip specifically enable, but do not require, use of two hands during dye injection and enable improved deployment of all fingers on one or both hands during the dye injection, enabling more force to be comfortably applied by the operator.
[0028] In an embodiment, (i) the top fingergrip is also placed such that the vertical plane of the rearmost interior wall of the finger ring is placed forward of the vertical plane of the rearmost interior wall of the interior bottom fingergrip, permitting more force to be exerted by the index finger during the injection, and (ii) the surface on which the third finger is placed is located forward of the surfaces on which the fourth and little fingers are placed, permitting more force to be exerted by the third finger during the injection. The barrel, nozzle and fingergrips are formed generally from a rigid material, and the nozzle, in an embodiment, may revolve around its longitudinal axis.
[0029] The plunger includes a plunger shaft, formed of a rigid material, and a plunger tip, formed of a generally resilient or elastomeric material. A barrel cap, located at the proximal end of the barrel, prevents the plunger from slipping out of the barrel when the plunger is moved proximally and may also be used to prevent the plunger from significantly rotating about its longitudinal axis.
[0030] The handgrip is located at the proximal end of the plunger and includes a handgrip base at the bottom, a handgrip arch in the middle, and a thumbrest and thumbhook at the top. The handgrip base extends significantly downward from the junction of the handgrip and the plunger, permitting a much greater surface area of the palm and ball of the thumb of the operator's hand to come in contact with the handgrip than in other manually-operated syringes. More particularly, the handgrip base extends, when being gripped by an operator, to the general area of the palmar surface of the annular ligament located at the base of the palm, and also is generally in contact with aspects of the flexor brevis pollicis, palmaris brevis, and abductor pollicis muscles of the hand; the handgrip arch is generally in contact with aspects of the adductor pollicis and immediately adjacent muscles of the hand.
[0031] The portion of the handgrip which includes the thumbrest and thumbhook extends upwards from the junction of the handgrip and plunger in generally the same axis as the other portions of the handgrip. The thumbrest enables use of the thumb on the hand opposite the hand gripping the interior bottom fingergrip to specifically enable, but not require, use of two hands during dye injection. The thumbhook provides a convenient means of moving the handgrip and plunger proximally, for purposes of drawing dye into the barrel from the manifold, using the thumb of the hand gripping the ergonomic syringe; alternatively, a finger from the opposite hand may be used to move the thumbhook. Additional features are included on the syringe to enhance its ergonomic performance, including surfaces on the handgrip and fingergrips sculpted to fit anatomical features on the operator's hand and fingers, and radii or bevels on corners and edges of gripping surfaces.
[0032] The ergonomic syringe therefore achieves its objects relative to the prior art, as follows:
[0033] (a) Increased comfort and reduced fatigue, strain and risk of disability compared to current syringes and the syringe described by Saez in U.S. Pat. No. 4,925,449 result from: (i) the flexibility of using either one or two hands to inject; (ii) the greatly increased surface area on the handgrip with which to exert force onto the plunger, and the shape and large surface area of the handgrip base, which distributes the pressure over a larger portion of the palm and ball of the thumb on the operator's hand, (iii) the locations of the fingergrips enable placement of fingers such that additional force may be applied without increasing operator discomfort. This object's achievement is particularly important as use of catheters sized 5 French and smaller increases.
[0034] (b) Easier injection of fluids during medical procedures is due to the increased comfort and reduced fatigue and strain for operators, and as a result of enabling one-handed operation for both loading the syringe with dye and injecting the dye; the thumbhook provides a convenient means of using the thumb on the hand gripping the syringe to draw back the syringe, permitting the other hand to be used for other purposes.
[0035] (c) Angiograms taken when the ergonomic syringe is used provide sufficient contrast because of the proper density of dye in the blood flow of the target vessels; this results from the operator being able to exert a larger amount of pressure to inject the dye without experiencing significant increase in discomfort, fatigue or strain compared to current manually-operated syringes and the syringe described by Saez in U.S. Pat. No. 4,925,449.
[0036] (d) The unit cost of the ergonomic syringe is expected to be similar to that of currently available syringes used for angiographic dye injections, less than that of the syringe described by Anwar in U.S. Pat. No. 6,030,368, and far less expensive than using electromechanical injectors, in particular because the cost of the injector equipment is very high.
[0037] (e) The ergonomic syringe functions similarly to other manually-operated syringes used for angiography, in that (i) force is exerted directly upon a plunger, (ii) the barrel is positioned between the index and third fingers, (iii) the syringe is handheld, and (iv) the connection to the manifold and operation of the plunger with respect to loading and injecting are the same as with manually-operated angiography syringes.
[0038]
[0039]
[0040] Fib. 2B is another back view of the invention.
[0041]
[0042]
[0043]
[0044]
[0045] An application of the ergonomic syringe is its use by an operator to inject dye into a patient during an angiographic procedure for the purpose of obtaining an angiogram showing the interior topography of blood vessels requiring evaluation. Although said ergonomic syringe may be used with a catheter of any size, it is particularly suited for use by an operator to inject dye through catheters sized 5 French or smaller, while increasing comfort and reducing the fatigue, strain, and risk of disability which may be associated with achieving the force required for such injections to achieve proper dye density in the blood flow through small catheters.
[0046] Referring to FIGS.
[0047]
[0048] A top fingergrip
[0049] The top fingergrip
[0050] The finger ring
[0051] The bottom fingergrip
[0052] The bottom fingergrip
[0053] The plunger
[0054] The handgrip
[0055] The thumbhook
[0056]
[0057]
[0058]
[0059]
[0060]
[0061] In another embodiment, the ergonomic syringe
[0062]
[0063] This detailed description of the invention is for illustrative purposes only. A reading by those skilled in the art will bring to mind various changes without departing from the spirit and scope of the invention.