Next Patent: Multiple focal spot X-ray inspection system
Next Patent: Multiple focal spot X-ray inspection system
Plaque It!
Sponsored by: Flash of Genius |
[0001] The present invention relates to x-rays production. More specifically, the present invention is concerned with a method and system for generating microfocused laser-based x-rays suitable for mammography
[0002] Mammographic signs of early cancer include calcifications, small masses (densities), neo-densities and architectural distortions. Screening mammography aims at revealing these signs and is highly sensitive and specific. However, still a notable fraction of mammography produces false positive or false negative results These outcomes most occur in the thick and/or dense breast. False positive results may result in an unnecessary additional imaging and biopsy. False negative results, which occur sometimes even when a cancer is palpable, have a severe adverse effect due to delayed biopsy and delayed cancer diagnosis.
[0003] Presence of calcifications usually helps to detect breast pathology. Although calcifications demonstrate well-pronounced contrast in mammograms, their detection threshold size is a function of x-ray tube focal spot size, radiographic technique factors including radiation quality, signal-to-noise ratio and exposure time, as well as function of detector and display devices. Even though, in early stages of breast disease microcalcifications are present in approximately half of the cases, they are not apparent on regular screening mammographs before reaching the detection threshold size, which is typically around 100 μμm. This especifically applies to punctuate calcifications
[0004] Some of the inherent limitations of the x-ray tube based screen-film mammography include:
[0005] relatively large focal spot (typically nominal ˜100 μm);
[0006] varying effective focal spot size across the imaged breast;
[0007] suboptimal (below 13 lp/mm) spatial resolution in the upper quadrants for breast thickness above 6 cm:
[0008] off-focal radiation;
[0009] a relatively low output (especially with microfocus);
[0010] a suboptimal spectral characteristic of x-rays for imaging dense fibroglandular tissue and/or thick breast;
[0011] restricted latitude of film-screen mammography, and
[0012] restricted contrast of film-screen mammography.
[0013] Regarding the two last restrictions, film-screen mammography utilizes film as a recording medium to properly record and display relatively narrow dynamic range of x-ray exposures. This should be contrasted with specially designed detectors that can correctly record four orders of magnitude of x-ray exposure. This information can be subsequently displayed on specially designed monitors. Moreover, since film-screen mammography aims at obtaining very high contrast images, the slope of the optical density/x-ray exposure curve is very high. As a result, only a very limited range of x-ray exposure is acceptable If exposure is too high or too low, it will produce exceedingly high or low optical density thus rendering the image not useful clinically.
[0014] While the two last restrictions can be alleviated to a great extent by digital mammography, the other limitations stem from the inherent limitations of x-ray tube and are unlikely to be overcome in the framework of this technology.
[0015] New technology for generating x-rays had emerged in recent years It relies on emission of x-rays from laser-produced plasma (LPP). This phenomenon occurs when a visible or infrared laser beam is focused onto the surface of solids or liquids If the optical power density exceeds a material dependent threshold value, continuous bremsstrahlung and characteristic x-ray emission lines result. However, initially very expensive and large laser systems were required to obtain the required optical power and the LPP x-ray sources were rather large (100 μm-1 mm). The invention of chirped-pulse amplification (CPA) in the late 1980's allowed achievement of high optical power density delivered to the target by the laser beam from compact and significantly cheaper table-top terawatt ultra-fast lasers.
[0016] The feasibility of CPA lasers for mammography and angiography has been demonstrated in both “High Magnification Imaging With a Laser-Based Hard X-ray Source”, IEEE Journal of selected topics in Quantum Electron, Special issue on laser in medicine, 5, 911-915 (1999), by J. Yu, Z. Jiang, J. C. Kieffer, A. Krol. and “Laser-Based microfocused X-ray Source for Mammography faisability Study”, Journal of Medical Physics, 24, 725-732 (1997), by A. Krol, A. Ikhlef, J C Kieffer, D. Bassano, C. C. Chamberlain, Z Jiang, H. Pepin, S. C. Parsad. Data in these publications shows the ability to obtain focal spot size of the order of 10 μm necessary to perform high spatial resolution mammographic imaging and confirm system ability to produce x-ray spectra from a number of different target materials, including Mo, Rh, Ag, In and, Sn with characteristic emission energies spanning 17 3 keV to 28.5 keV.
[0017] Even though LLP x-ray source created by CPA lasers can be very small (10 μm or less) and bright, with peak power many order of magnitude higher than conventional x-ray tubes, a drawback of CPA laser sources from the prior art has been their low average power However, in recent years, significant progress has been made in this respect 20 W average power CPA lasers are presently available, and a 50 W CPA laser is in the design stage
[0018] Finally, there have been studies on sub-picosecond laser-solid-matter interaction devoted to investigation of x-rays generation, from the soft x-rays up to the very hard x-ray emission in the MeV range. However, nobody attempted to simultaneously control the emitted x-ray spectrum, the x-ray source size and conversion efficiency from laser to x-rays.
[0019] Method and system for generating microfocused laser-based x-rays allowing to simultaneously optimize the x-ray source size, its spectral distribution, and the conversion efficiency in the 17 3-28.5 keV range is thus desirable.
[0020] An object of the present invention is therefore to provide an improved system and a method for generating microfocused laser-based x-rays.
[0021] Specifically an object of the present invention is to provide such method and system allowing optimizing the x-ray source size, its spectral distribution, and the conversion efficiency in the 17.3-28 5 keV range (adapted to the breast thickness and/or density) for improved mammography.
[0022] More specifically, in accordance with an aspect of the present invention, there is provided a system for generating microfocused laser-based x-rays comprising an ultra-fast laser; and a solid target with elemental composition positioned in the focal plane of said ultra-fast laser; wherein said ultra-fast laser and solid target are configured so that interaction with said solid target of a laser beam produced by said ultra-fast laser produces suprathermal electrons that penetrate said solid target in a spot whose size is comparable to the ultra-fast laser focal spot size, yielding hard x-rays More specifically, in accordance with another aspect of the present invention, there is provided
[0023] A method for generating microfocused laser-based x-rays using the above system, the method comprising: without The solid target, generating and amplifying from the ultra-fast laser, a first laser beam pulse train so as to control and adjust the laser energy, positioning the target in the focal plane of the ultra-fast laser so as to offer a fresh surface to the ultra-fast laser; generating and amplifying from the ultra-fast laser, a second laser beam pulse train, and releasing and focusing the second laser beam pulse train on the solid target; whereby the interaction of the second laser beam pulse train on the target generates hard x-rays.
[0024] A system and method according to the present Invention allows producing a very small x-ray source with an x-ray spectrum tailored to a specific patient's breast density and thickness suitable to detect pathology in examined breast, for example microcalcifications, with increased sensitivity and specificity, as compared to method and system from the prior-art.
[0025] Compared to systems and methods from the prior art, a system and method according to the present invention allows to lower calcification detection threshold size down to approximately 30 μm, improves visualization of morphology of microcalcifications and marginal characteristics of masses, improves detection of densities (masses), especially in a dense breast tissue, lowers mean grandular dose, especially in a dense and thick breast tissue, and enhances both sensitivity and specificity of mammography.
[0026] A method and system according to the present invention yields very small (below 30 μm) x-ray focal spot size that will results in:
[0027] significantly improved and uniform within field of view spatial resolution of the order of 20 lp/mm. In conventional mammography resolution is non-uniform and do not exceed 13 lp/mm. The upper quadrants of thicker (d>6 cm) breasts are imaged with resolution below 10 lp/mm and there is strong gradient of spatial resolution across field of view (in the cathode-anode direction) in the detector plane,
[0028] better visualization of pathologies, including demonstration of morphology of very small microcalcifications that cannot be imaged with conventional mammography, that will allow earlier detection of breast cancer with significant benefit for women's health; and
[0029] magnification imaging without necessity of moving breast closer to the source This is necessary in conventional mammography and results in decreased field of view
[0030] Moreover a method and system according to the present invention allows x-ray spectrum to be tailored to specific breast thickness/density that will result in:
[0031] lower dose to thicker/denser breast with better contrast, as compared to conventional mammography. Dose saving can be as large as factor of three for very thick/dense breast (d>>4.5 cm); and
[0032] better imaging of thicker/denser breast due to shorter imaging time. It might completely eliminate “white mammograms” presently obtained with radiopaque breast.
[0033] Other objects, advantages and features of the present invention will become more apparent upon reading of the following non-restrictive description of preferred embodiments thereof, given by way of example only with reference to the accompanying drawings.
[0034] In the appended drawings:
[0035]
[0036]
[0037]
[0038]
[0039]
[0040]
[0041] Turning to
[0042] The x-ray generator
[0043] As it will be explained hereinbelow in more detail, the ultra-fast laser
[0044] The target
[0045] The target/filter allows to properly shape the x-ray spectrum A precision target positioning/feedback allows to keep the x-ray focal spot small.
[0046] The precision target positioning system
[0047] The x-ray window
[0048] Of course, the target chamber
[0049] The x-ray generator
[0050] The optical beam transfer system
[0051] Of course, the optical beam transfer system
[0052] The target chamber
[0053] The x-ray mammography system
[0054] The compression paddles
[0055] The image receptor
[0056] Of course, the x-ray mammography system
[0057] The operation of the x-ray generator
[0058] Without target
[0059] The target
[0060] It is to be noted that the motion of the target (rotation and or translation in a fixed plane) is triggered in order to offer a fresh surface at every laser shot.
[0061] A second sequence of high voltages is then triggered for the generation and amplification of the interacting laser beam pulse train with appropriate energy and pulse number. The pulse number is selected using an adjustable gate
[0062] The pulse is compressed in an optical compressor composed of a set of gratings and mirrors. The shutter is opened at the compressor exit to release the pulse train on the target
[0063] Hard x-rays
[0064] Turning now to
[0065]
[0066] The angle of incidence of the laser pulse
[0067] where T
[0068] A judicious choice of target geometry and composition, of the laser pulse duration, polarization and/or intensity, allows shaping the x-ray spectrum and to maintain and control the hot electron temperature between 10 and 50 keV. The conversion efficiency is drastically reduced for hot electron temperatures lower than 10 keV It has also been found that hot electron temperatures higher than 50 keV induce a degradation of imaging contrast due to the high-energy tail in the x-ray spectrum The emitted x-ray spectrum is composed of continuum bremsstrahlung and discrete emission lines. The K
[0069] As it will be explained hereinbelow in more detail, the laser
[0070] According to the present invention, the radiation pressure of the pulse produced by the laser
[0071] The laser parameters are controlled as follows. A very high contrast laser pulse (10
[0072] The control of the x-ray source size imposes utilization of a well-defined narrow intensity range (10
[0073] The hot electron temperature can be adjusted by a careful laser intensity control indeed, it has been found through experiments that
[0074] where l is the laser intensity and λ is the laser wavelength. Radiation Confined Plasmas yields α=5×10
[0075] This optimum intensity as a function of the target material atomic number Z is shown in
[0076] The laser intensity window, for example, for a laser having a 400 fs pulse and 0 53 μm wavelength, is in the 2×10
[0077] using RCP and thus controlling the hard x-ray source size;
[0078] utilizing targets with the appropriate elemental composition to match the x-ray energy to a patient specific breast thickness/density with an optimized conversion efficiency (η
[0079] It has been found that a pulse duration between 200 fs and 600 fs allows to optimize the plasma confinement effect Through radiation pressure and thus to optimize the x-ray source parameters. Through experiments and calculations, it has also been found that shorter (150 fs) and longer (1 ps) pulse resulted in lower conversion efficiency and/or larger x-ray spot size. According to a preferred embodiment of the present invention, a typical value used for the pulse duration is 400 fs (with no pre-pulse and/or pedestal)
[0080] A theoretical investigation of the optimum monochromatic x-ray energy for imaging various simulated pathologies in the breast for a large range of breast thickness spanning the 2-cm-8-cm range was performed. The following figure of merit (FOM) was evaluated
[0081] Simulated pathologies included 200-μm calcifications and 5-mm masses.
[0082] It has been confirmed that the shape and the exact location of maximum depend on the imaging object. However in each case the optimum energy shifts towards higher energy for thicker/denser breast for all the investigated pathologies. It has been found that the optimum photon energy increases from ˜15-18 keV for 2 cm Lucite equivalent breast to ˜25-28 keV for 8 cm Lucite equivalent breast.
[0083] For larger breast thickness/density (d>4.5 cm) imaged with ultra-fast-based laser x-ray source, we expect significant improvement in the dose utilization without loss of contrast, as compared to imaging with a mammographic x-ray tube operating at increased kilovoltage (28-31 kVp)
[0084] The target
[0085] Preferably, an x-ray generator
[0086] In both cases, hot electron
[0087] As illustrated in
[0088] The operation of the overall x-ray mammography system
[0089] Although the present invention has been described hereinabove by way of preferred embodiments thereof, it can be modified, without departing from the spirit and nature of the subject invention as defined in the appended claims.