Plaque It!
Sponsored by: Flash of Genius |
[0001] This invention relates to mass spectrometers. More particularly, this invention relates to tandem mass spectrometers, intended to perform multiple mass analysis or selection steps.
[0002] Presently, a variety of mass spectrometry/mass spectrometry (MS/MS or MS
[0003] Triple quadrupole mass spectrometers (TQMS) accomplish these steps through the use of two quadrupole mass analyzers separated by a pressurized reaction region for the fragmentation step. Since the three steps of the MS/MS process are carried out in different locations, MS/MS using a triple quadrupole mass spectrometer is referred to as “tandem in space”. MS/MS spectra with a TQMS can be quite complex in terms of the number of mass resolved features due to the tens of electron volts laboratory collision energies used and the fact that once a fragment ion is formed it can undergo further decomposition producing additional second generation ions and so on. The resulting MS/MS spectrum is a composite of all the fragmentation processes that are energetically allowed: precursor ion to fragment ions and fragment ions to other fragment ions. This spectral richness is often a benefit to compound identification when searching databases of MS/MS libraries. However, this same spectral complexity can make structural identification of a completely unknown compound difficult since not all of the fragment ions in the spectrum are first generation products from the precursor ion.
[0004] There are also situations in which the MS/MS spectrum yields only one or two fragment ion features that correspond to loss of a structurally insignificant part of the precursor ion. The data from these MS/MS spectra are not particularly helpful for determining the structure of unknown precursor ions.
[0005] An additional stage of MS applied to the MS/MS scheme outlined above, giving MS/MS/MS or MS
[0006] Three-dimensional ion traps provide the capability of multiple stages of MS/MS (often referred to as MS
[0007] True MS
[0008] An additional stage of fragmentation within a flow-through pressurized collision cell, but without the isolation step has been demonstrated for a QqTOF instrument as described by Cousins [47th ASMS Conference on Mass Spectrometry and Allied Topics, 1999]. Here, a precursor ion is selected within the first quadrupole mass analyzer, and then accelerated into the collision cell where primary fragment ions are produced. Further fragmentation of a selected primary fragmentation is induced by an appropriately chosen AC voltage source that is resonant with the particular, primary, fragment ion. This excited primary fragment ion then undergoes further collisions with background neutral species and dissociates, to generate secondary fragment ions. The result is a MS
[0009] An alternative approach is to trap the ions within the collision cell and this offers the opportunity to both isolate and fragment a chosen ion using techniques analogous to those used in a conventional three-dimensional ion trap. Theoretically, this should overcome the flow through characteristics, resulting in insufficient time for additional fragmentation, noted above. The problem with this approach is that once the ions are released from the collision cell the downstream mass spectrometer must perform the mass analysis step very quickly since the pulse of released ions is temporally very narrow. This requires that the downstream mass analyzer be a very fast scanning device, such as a TOF mass spectrometer.
[0010] Thus, a conventional scanning quadrupole mass analyzer or the like is unsuited for processing a temporally narrow pulse of ions. If the ions could somehow be scanned out of the trap in some mass-dependent manner, this difficulty could be overcome.
[0011] In earlier U.S. Pat. No. 6,177,668, also published international application WO 97/4702, there is disclosed a multipole mass spectrometer provided with ion trap and an axial ejection technique from the ion trap. The contents of these two applications are hereby incorporated by reference.
[0012] The technique disclosed in those two applications, relies upon emitting ions into the entrance of a rod set, for example a quadrupole rod set, and trapping the ions at the far end by producing a barrier field at an exit member. An RF field is applied to the rods, at least adjacent to the barrier member, and the RF fields interact in an extraction region adjacent to the exit end of the rod set and the barrier member, to produce a fringing field. Ions in the extraction region are energized to eject, mass selectively, at least some ions of a selected mass-to-charge ratio axially from the rod set and past the barrier field. The ejected ions can then be detected. Various techniques are taught for ejecting the ions axially, namely scanning an auxiliary AC field applied to the end lens or barrier, scanning the RF voltage applied to the rod set while applying a fixed frequency auxiliary voltage to the end barrier and applying an auxiliary AC voltage to the rod set in addition to that on the lens and the RF on the rods.
[0013] It has now been realized that this 2-dimensional linear ion trap mass spectrometer can be used to enhance the performance of a triple quadrupole to provide MS
[0014] In accordance with a first aspect of the present invention, there is provided a method of analyzing a substance, the method comprising:
[0015] (1) ionizing the substance to form a stream of ions;
[0016] (2) subjecting the ions stream to a first mass analysis, to select ions having a desired mass to charge ratio, as precursor ions;
[0017] (3) introducing the precursor ions into a collision cell to promote fragmentation of the precursor ions, thereby to generate primary fragment ions;
[0018] (4) in the collision cell, selecting primary fragment ions having a desired mass to charge ratio, and rejecting other ions;
[0019] (5) accelerating the selected primary fragment ions from the collision cell into a downstream mass analyzer, thereby to promote secondary fragmentation; and
[0020] (6) mass analyzing the secondary fragment ions to generate a mass spectrum.
[0021] In accordance with a second aspect of the present invention, there is provided a method of analyzing a substance, the method comprising:
[0022] (1) ionizing the substance to form a stream of ions;
[0023] (2) subjecting the ions stream to a first mass analysis, to select ions having a desired mass to charge ratio, as precursor ions;
[0024] (3) introducing the precursor ions into a collision cell to promote fragmentation of the precursor ions, thereby to generate primary fragment ions;
[0025] (4) in the collision cell, selecting primary fragment ions having a desired mass to charge ratio, and rejecting other ions by removing ions of a mass to charge ratio greater than the mass to charge ratio of the selected primary fragment ions and separately removing ions with a mass to charge ratio less than the mass to charge ratio of the selected primary fragment ion, the removal of the ions with mass to charge ratios higher and lower than the mass to charge ratio of the selected primary fragment ion being effected in either order;
[0026] (5) causing the selected primary fragment ions to collide, to promote further fragmentation, generating secondary fragment ions; and
[0027] (6) mass analyzing the secondary fragment ions to generate a mass spectrum.
[0028] For a better understanding of the present invention and to show more clearly how it may be carried into effect, reference will now be made, by way of example, to the accompanying drawings which show a preferred embodiment of the present invention and in which:
[0029]
[0030]
[0031]
[0032]
[0033]
[0034]
[0035]
[0036]
[0037]
[0038]
[0039]
[0040]
[0041]
[0042]
[0043]
[0044]
[0045]
[0046] Referring first to
[0047] The ions then pass through an orifice
[0048] In the chamber
[0049] An interquad aperture IQ1 separates the chamber
[0050] Power supplies
[0051] In the preferred embodiment, ions from ion source
[0052] The subsequent collisional activation step can be accomplished as in a conventional three-dimensional ion trap, that is by application of an appropriate resonant AC waveform. This however requires sophisticated electronics and has the additional requirement that the trapping RF voltage be such that the lowest mass fragment ion and the precursor ion are simultaneously stable within Q2.
[0053] An alternative technique is to simply accelerate the mass isolated ions in to the subsequent mass analyzer. Since Q2 is operated at elevated neutral gas pressure, say 5×10
[0054] By way of example consider the following set of experimental results obtained using the apparatus in
[0055] Ion isolation of the m/z 397 fragment ion was accomplished in a step-wise fashion by first adjusting the RF voltage applied to the Q2 rods such that ions above m/z 397 become unstable within Q2 and are lost. The result of this step is displayed in
[0056] Low mass ions may be eliminated from the Q2 ion population by adjusting the RF voltage such that the trapped ions with m/z below 397 become unstable in the Q2 and are also lost. The result of this step prior to mass analysis is displayed in
[0057] A combination of these two steps thus provides good mass isolation of the m/z 397 fragment ion within Q2 as is displayed in
[0058] Finally, the m/z 397 ions are accelerated into the Q3 linear ion trap MS by increasing the relative DC voltage offset between Q2 and Q3 from 5 volts (used in
[0059] This procedure can be carried out separately on the major fragment ions in the reference reserpine MS
[0060] For this technique to be widely applicable the collisional activation step must be sufficiently energetic to provide a wide range of MS
[0061]
[0062] It is understood that the ion isolation step can be accomplished via notched broadband isolation techniques. This entails subjecting the trapped ions to a plurality of excitation signals uniformly spaced in the frequency domain with a notch of no excitation signals corresponding to the resonant frequencies of the ions to be isolated within the ion trap as described by Douglas et al. in WO 00/33350.
[0063] The present inventors have also discovered and identified that one of the important experimental parameters in the transfer of ions from the Q2 linear ion trap to the Q3 linear ion trap is the RF voltage value applied to the Q3 linear ion trap during the Q2-to-Q3 ion acceleration process. Ions received in Q3 can only be successfully trapped within Q3 if their associated q-value is less than ˜0.9.
[0064] The inventors have found that another important parameter is the time for which the Q3 RF voltage is held at the fill mass, referred to as the “Q3 fill time”. This Q3 fill time is in general longer than the actual time required to empty the Q2 ion trap. Ions can be removed from Q2 very rapidly by using an axial DC field as taught by Thomson and Jolliffe in U.S. Pat. No. 6,111,250. At the pressures and voltages used in the current instrument all the ions within Q2 should be transferred to the Q3 ion trap in less than 2 ms, which can be identified as a “transfer time”. Any time in excess of this 2 ms or other transfer time but less than the Q3 fill time is referred to as the “delay time”.
[0065] The Q3 fill time for the experiment that resulted in the spectrum displayed in
[0066] It is necessary to consider the details of the scanning procedure to understand the reason for the appearance of the low mass-to-charge product ions in the
[0067] As indicated at 30, at the end of the fill time, the voltage on the interquad aperture IQ3 is increased to a potential indicated at 32. Simultaneously, the voltage on the exit lens
[0068] At the end of the Q3 fill time, the voltage on the exit lens
[0069] The inventors have found that a very important factor influencing whether or not ions with mass-to-charge ratios below that of the Q3 fill mass are observed is the duration of the Q3 fill step, i.e. the Q3 fill time up to the voltage changes indicated at 26 and 30 in
[0070] It is believed that the reason for the observation of ions with q-values seemingly greater than the first stability region limit of ˜0.908 is the unique Q2-to-Q3 fragmentation environment. The pulse of ions was introduced into the Q3 linear ion trap at a translational energy of 35 eV
[0071] One advantage to operating the instrument with a high Q3 fill mass is a higher intensity product ion mass spectrum relative to that obtained with a low Q3 fill mass.
[0072] A further advantage to the use of an elevated Q3 fill mass is that the ions with m/z<Q3 fill mass are produced at a later time (after the cooling time) than those with m/z>Q3 fill mass, as they are products of precursor ions with lower kinetic energy since some collisional relaxation of the precursor ion during the delay time. That is, the energy of the precursor ion has been reduced by some of the relatively infrequent collisions within Q3 during the fill time. Thus consecutive fragmentation processes producing these ions with m/z<Q3 fill mass are less favoured since the precursor ion has less internal energy at the time at which the lower mass product ions are collected. The resulting product ions in turn have less internal energy and thus reduced probability of further fragmentation, leading to suppression of second generation product ion precursor-to-product ion pairs. This can make it easier to identify first generation precursor-to-product ion pairs, which can be especially useful in the identification and differentiation of different dissociation pathways.
[0073] An example is the mapping of the product ions of bosentan studied by Hopfgartner et. al. (J. Mass Spectrom.
[0074] A product ion mass spectrum for bosentan was obtained using the method described herein. Once again the precursor ion was mass selected by Q1 and then, in accordance with the present invention, it was introduced into and trapped within Q2, this time at low energy in order to eliminate fragmentation. Next, the ions trapped within Q2 were accelerated into the Q3 linear ion trap at a laboratory collision energy of 30 eV, a Q3 fill mass of 400, and a Q3 fill time of 5 ms (i.e. 2 ms transfer time and 3 ms delay time). Thus, the only product ions that would be stable during the 5 ms fill time in the Q3 ion trap have m/z>400. Immediately after the Q3 fill time (at 26 in
[0075] The product ion mass spectrum of the m/z 552 bosentan molecular ion obtained with the Q3 fill mass set at 400 for a 10 ms fill time (i.e. 2 ms transfer time and 8 ms delay time) is displayed in
[0076] The only limitation for the use of a variable Q3 fill mass is that the precursor ion must be stable within the Q3 linear ion trap, so the Q3 fill mass must be less than the mass-to-charge ratio of the precursor ion.
[0077] This method has also been found to be useful for the simplification of peptide product ion spectra as is demonstrated in
[0078] The technique of ion isolation within a nominally RF-only collision cell and subsequent ion acceleration with concomitant fragmentation is also applicable to other Qq(MS) (where Q designates the mass selection step via a conventional RF/DC resolving quadrupole mass spectrometer and q the higher pressure nominally RF-only collision cell , here carried out in Q1 and Q2 respectively) instruments, where the MS stage can be another fast scanning mass spectrometer other than a linear ion trap mass spectrometer. One such device is a QqTOF tandem mass spectrometer. The TOF is particularly well suited to be used for the final mass analyzer since it is best used with a pulsed ion source, which is what emerges from the collision cell. Furthermore, a full mass spectrum can be obtained for each ion pulse, giving better overall efficiency.
[0079] Additionally, it may in some circumstances be possible to eliminate the collision cell, and provide the collision gas by some other mechanism to the flow of ions into Q3. Additionally, the basic requirement for the section of containing Q3 is that it will be a lower pressure section capable of collecting and collimating ions. It could include, for example, a multipole rod set that provides just this function without acting as a mass analyzer. Where it is desired to set a fill mass, the multipole rod set must be capable of defining this cut off mass with a required degree of precision. A mass analyzer can then be provided downstream.
[0080] The final step of mass analyzing the MS